Comparative Review of Blood Cell and Immunoassay Analyzers in Sports‑Specific Biochemical Laboratories
Sports biochemical laboratories utilize athlete blood samples to conduct physiological assessments, primarily employing blood cell analyzers and immunoassay analyzers. This study systematically reviews the progress and limitations of these two instrument types in this field. A comprehensive literature review shows that comparative studies involving blood cell analyzers have a well-established practical foundation, and results show good comparability between instruments; however, more precise comparative protocols for athlete blood samples characterized by median concentrations remain lacking. In contrast, comparative studies on immunoassay analyzers have predominantly focused on medical laboratories, while key indicators for physiological assessment in athletes have not yet received sufficient attention. The lack of cross-system consistency in immunoassay analyzers remains a critical bottleneck limiting the comparability of longitudinal monitoring data in athletes. For future research, sports biochemical laboratories are advised to: ① establish standardized instrument comparison protocols specifically tailored to athlete median-concentration blood samples; ② prioritize investigation of systematic biases in key hormonal markers—such as testosterone and cortisol—across different immunoassay platforms, and assess their impact on longitudinal physiological assessments; ③ explore the development of specialized instrument comparison quality control systems for sports biochemical laboratories.
[1] Yu WH, Wu HX, Mo QN, et al., 2017, Validation of Comparability and Bias Assessment of Test Results from Different Testing Systems Within the Same Medical Institution. Chinese Journal of Clinicians (Electronic Edition), 11(6): 931–937.
[2] Ma XY, 2026, Application of the Mindray BC-7500CS Hematology Analyzer in Body Fluid Cell Counting and Differential Counting. China Medical Device Information, 32(4): 30–32.
[3] Yao LT, Ye XJ, 2011, Performance Evaluation and Result Comparison of Two Hematology Analyzers. International Journal of Laboratory Medicine, 32(19): 2255–2257.
[4] Xu JY, 2019, Comparative Study on Two Different Models of Blood Cell Analyzers. Smart Health, 5(22): 7–8.
[5] Song AJ, Qiu W, Zhang Y, 2017, Comparative Study on Test Results from Different Blood Cell Analyzers in a Clinical Biochemistry Laboratory in Sports Medicine Settings. International Journal of Laboratory Medicine, 38(18): 2534–2536.
[6] Pan XY, Hu HM, 2018, Performance Evaluation and Optimization Recommendations for Medical Testing and Inspection Systems. Journal of Clinical Laboratory Medicine (Electronic Edition), 7(3): 567.
[7] Zhang K, Guo XZ, Sun DD, et al., 2017, Validation and Comparative Evaluation of the Analytical Performance of Beckman Coulter 4th and 5th Generation Total β-Subunit Human Chorionic Gonadotropin Assay Reagents. Lab. Med. Clin. Anal., 24(2): 221–225.
[8] Ficarra G, Caccamo D, Rottura M, et al., 2023, Testosterone: Cortisol Ratio as a Predictor of Podium in Adolescent Rowing Athletes. Heliyon, 9(11): e22315.
[9] Cupka M, Sedliak M, 2023, Hungry Runners – Low Energy Availability in Male Endurance Athletes and Its Impact on Performance and Testosterone: Mini-Review. European Journal of Translational Myology, 33(2): 11104.
[10] Roli L, De Vincentis S, Rocchi MBL, et al., 2018, Testosterone, Cortisol, hGH, and IGF-1 Levels in an Italian Female Elite Volleyball Team. Health Science Reports, 1(4): e32.
[11] Gandouzi I, Turki M, Ayedi F, et al., 2021, Testosterone and Cortisol Variation Due to Training and Fight in Judokas. Journal of Sports Medicine and Physical Fitness, 61(12): 1690–1699.
[12] International Organization for Standardization, 2022, ISO 15189:2022 Medical Laboratories – Requirements for Quality and Competence. International Organization for Standardization.
[13] National Accreditation Service for Conformity Assessment, China, 2023, CNAS-CL02:2023 Accreditation Criteria for Quality and Competence of Medical Laboratories. Beijing: National Accreditation Service for Conformity Assessment, China.
[14] CLSI, 2010, EP9-A2 Method Comparison and Bias Estimation Using Patient Samples. Wayne, PA, USA: Clinical and Laboratory Standards Institute.
[15] CLSI, 2013, EP09-A3 Measurement Procedure Comparison and Bias Estimation Using Patient Samples. Wayne, PA, USA: Clinical and Laboratory Standards Institute.
[16] Ministry of Health of the People's Republic of China, 2013, WS/T 406-2012 Quality Requirements for Analysis of Routine Clinical Hematology Test Items. Beijing: China Standards Press.
[17] Ministry of Health of the People's Republic of China, 2013, WS/T 407-2012 Quality Requirements for Analysis of Clinical Quantitative Testing Items. Beijing: China Standards Press.
[18] Zhuang JH, Huang XZ, Xu N, et al., 2023, Part V: Examples of Quality Management for Testing Processes. In: Zhuang JH, Huang XZ, Xu N, Ke PF, eds. Examples of Quality System Documents for Medical Laboratories. Beijing: People's Medical Publishing House, 714–823.
[19] An CW, Chen J, Pang L, et al., 2022, Methodological Comparison and Bias Assessment of Two Procalcitonin Detection Systems Using the CLSI EP9-A2-IR Document. Lab. Med. Clin. Pathol., 29(7): 1224–1228.
[20] Ren J, Wang WJ, Wang W, et al., 2026, Performance Validation of Coagulation Analyzers Based on the New Standard WS/T406-2024. Journal of Hebei Medical University, 47(6): 710–716.
[21] Wu HN, Xiong YT, 2026, Overview of Applicability Evaluation of Comparable Sample Types in Quantitative Detection Reagents. Journal of Molecular Diagnostics and Therapy, 18(1): 222–224.
[22] Xie XB, 2025, Chapter 3: ISO 15189 Knowledge Question Bank. In: Xie XB, Chen QB, Wang LX, et al., eds. Basic Knowledge of Medical Laboratory Accreditation and Analysis of Typical Nonconformities. Beijing: People's Medical Publishing House, 251–321.
[23] Liu H, 2018, Comparative Analysis of Test Results from Different Blood Cell Analyzers. Sichuan Journal of Anatomy, 26(2): 41–42.
[24] Kang C, Zhang X, Huang LM, et al., 2021, Comparative Evaluation of Clinical Test Results Between Different Models of XiSenMeiKang Complete Blood Count Analyzers. China Medical Device Information, 27(4): 43–44.
[25] Zuo JH, Liu XR, Cheng SL, et al., 2019, Evaluation of Test Result Consistency Among Three Different Models of Photoelectric Blood Cell Analyzers. Journal of Laboratory Medicine and Clinical Practice, 16(6): 799–801.
[26] Luo M, Feng PF, Zhang YJ, et al., 2025, Performance Validation of Sex Hormone Detection Using the Mindray CL6000i Chemiluminescence Immunoassay Analyzer. China Medical Device Information, 31(17): 53–58.
[27] Han YC, Zheng Y, Mao J, et al., 2025, Evaluation of Two Chemiluminescence Immunoassay Analyzers for Serum VitB12 Detection. Journal of Binzhou Medical University, 48(2): 141–144.
[28] Li XY, Zhou XY, Li FQ, 2023, Comparative Analysis of Serum β-HCG and Progesterone Detection Using the Cobas e 411 Electrochemiluminescence Fully Automated Immunoassay Analyzer and the Beckman UniCel DxI 800 Chemiluminescence Analyzer. China Medical Device Information, 29(9): 120–122.
[29] Sun HZ, Dong J, 2024, Exploration of a Range Comparison Method Based on Clinical Significance for Different Chemiluminescence Immunoassay Analyzers. China Medical Device Information, 30(14): 39–42.
[30] Tan YG, Liu N, Tian Y, et al., 2017, Current Status Analysis of Determining Serum Human Growth Hormone Levels Using Various Chemiluminescence Assay Systems. Journal of Laboratory Medicine and Clinical Practice, 14(2): 161–163.
[31] Liu Z, 2023, Application Study of Convergence Transformation Methods for Test Results Across Different Instruments, thesis, Sichuan University.
[32] Fan R, Aynur A, Cao XQ, et al., 2021, Food Microbial Testing and Laboratory Quality Control. Journal of Food Safety and Quality Inspection, 12(9): 3824–3829.
[33] Feng CY, Fu CD, Li JL, 2020, A Brief Discussion on Quality Management in Microbial Laboratories of the Food and Drug Regulatory System. Food Safety Guide, (12): 61–62.
[34] Zheng JJ, Ma JB, Wang CY, et al., 2020, POCT Quality Refined Management and Practice. Zhejiang University Press.
[35] Kumar BR, Férec C, Raut S, 2023, A Comparative Evaluation of Performance of Sysmex XN 3000 and Horiba Yumizen H2500 Automated Complete Blood Count Analyzers. Indian Journal of Hematology and Blood Transfusion, 40(2): 303–314.
[36] Gupta SG, Goyal D, Patel CG, et al., 2021, Comparison of Different Autoanalyzers for the Determination of Lymphocyte and Neutrophil Counts in Mouse Blood. Drug Metabolism and Personalized Therapy, 37(2): 219–222.
[37] Laaksonen AM, Kurkela K, Kuitunen K, et al., 2022, HemoScreen Hematology Analyzer Compared to Sysmex XN for Complete Blood Count, White Blood Cell Differential, and Detection of Leukocyte Abnormalities. eJHaem, 3(4): 1126–1134.
[38] Solis RCIJ, Laguna J, Julian J, et al., 2026, Comparison of Haemoglobin Assessments with Hemo Control and Sysmex XN-1000™ in High Altitude Resident Patients (Huancayo – Peru). Semergen, 52(2): 102702.
[39] Sharma VK, Nair A, Thomas ET, et al., 2025, Tacrolimus (FK-506) Levels Determined by 2 Different Chemiluminescence Immunoassay-Based Analytical Methods: A Comparative Cross-Sectional Study. Medicine (Baltimore), 104(39): e44895.
[40] Agrawal N, Sinha A, Chakraborty PP, et al., 2025, A Cross-Sectional Study on Comparison of Serum Cortisol Concentration Measured by Chemiluminescent Immunoassay in Four Different Automated Analyzers in a Variety of Adrenal Disorders. Indian Journal of Endocrinology and Metabolism, 29(1): 83–88.
[41] Medrano DSIJ, Laguna J, Julian J, et al., 2025, Comparison of Two Automated Immunoassays for Quantifying ProGRP, SCC and HE4 in Serum: Impact on Diagnostic Accuracy. Scandinavian Journal of Clinical and Laboratory Investigation, 85(2): 1–9.
[42] Yeoh AC, Ramdzan K, Jamaluddin FA, et al., 2024, B-032 Evaluation of Assay Performance for Atellica CI 1900 Analyzer in Routine Chemistry, Division of Laboratory Medicine, UMMC. Clinical Chemistry, 70(Suppl 1): 396.